3.7.5 Wipe Turbidity Sensor
When taking turbidity measurements or using the PurgeScan stabilization function on the MP25T system, if you
notice that turbidity values are rising significantly but the water in the flow cell does not appear to be increasing
in turbidity, bubbles forming from dissolved gases in the groundwater may be interfering with the sensor. When
this button is pressed, the wiper located on the end of the sensor will complete one “Wipe Cycle” to clean the
optical sensor surface of any debris or bubbles that can cause falsely high turbidity readings. Note that if this
button is pressed when connected to an MP25, nothing will happen.