Q-SENSE E4 LIQUID HANDLING SET UP

 

Q-Sense E4 Flow System Setup, with IPC-N Pump 

 
Part 1 – Ismatec IPC-N Pump 

 

Before switching the pump on, 
check the voltage setting and fuse 
on the backside: 
 
220-240V: use 250mA fuses 
110-120V: use 500mA fuses 

                           

                

 

    

 

 

 

Keep the “SETTINGS” button 
pressed while switching the pump 
on. This makes the pump start in 
settings mode (the blue words 
above the buttons are active). 

          

 

 

Cycle through the settings, using 
the up/down arrows, until “TUBE” is 
shown, select by pressing “OK”. 
 
Select inner diameter 

0.64

 mm, 

(double-check with numbers given 
on the tubing itself!). Press “OK” to 
confirm. 
 
Press “ESCAPE” to save settings. 

 
 
 
  

 

 

Attach the adaptors on each 
cassette, and then the tubing. 
 
Ready to go! 
 
 
 
 

 
 
 
 
 
 

 

 

Typical flow rates: for filling the 
system 100-400 

μ

l/min; for 

experiment 25-100 

μ

l/min. 

 
 

                                                                     

50  

μ

l/min 

 

 
                                            

100 

μ

l/min

 

 

 

 

For details and calibration function, see supplier’s manual.

 

 

 