Table 3.
Test Metal Electrode Default Settings
Common Designation
UNS
Alloy Factor
Default 4–20mA
General
Corrosion Scale
(MPY)
Default 4–20mA
Localized
Corrosion Scale
(Index)
Aluminum AA1100
A91100
0.94
0–50
0–100
Aluminum AA6061
A96061
0.94
0–50
0–100
Aluminum AA2024
A92024
0.86
0–50
0–100
Cu/Ni - 70/30
C71500
1.50
0–5
0–10
Copper 110 ETP
C11000
2.00
0–5
0–10
CDA 687 Aluminum
Brass Arsenical
C68700
1.62
0–5
0–10
CDA 642 Aluminum
Silicon Bronze
C64200
1.48
0–5
0–10
Arsenical Admiralty
Brass CDA443
C44300
1.67
0–5
0–10
Phosphorized Admiralty
Brass CD445
C44500
1.68
0–5
0–10
Pipe Grade Carbon Steel
A135
1.00
0–50
0–100
Mild Steel C1010
G10100
1.00
0–50
0–100
Mild Steel C1015
G10150
1.00
0–50
0–100
Mild Steel C1018-C1020
G10180
1.00
0–50
0–100
Mild Steel C1080
G10800
1.00
0–50
0–100
Stainless Steel 304
S30400
0.89
0–2.5
0–10
Stainless Steel 304L
S30403
0.89
0–2.5
0–10
Stainless Steel 316
S31600
0.90
0–2.5
0–10
Stainless Steel 316L
S31603
0.90
0–2.5
0–10
Duplex Stainless 2205 -
F51
S31803
0.90
0–2.5
0–10
Duplex Stainless 2507 -
F53
S32750
0.90
0–2.5
0–10
Common Lead
L50045
2.57
0–2.5
0–1
*
NOTE
*
Metal electrodes with a different length and a slightly different diameter may be
used with the CR-301 sensor if the electrodes have a 4-40 internal thread and has a surface
are of 5 cm
2
. The alloy factor for a given metal alloy is proportional to the surface area of
the electrode. If an electrode with a surface area different than 5 cm
2
(0.736 square inch)
is used, the user will need to calculate the proper alloy factor according to the surface area
and the default 5cm
2
alloy factor listed above.