2
Specifications
Table 1.
ST-600 Series Specifications
Specification
*
ST-600
ST-604
Part Number (P/N)
50231
50219
Sodium Hypochlorite
Range
0–16% mass/mass
(RTD Compensated)
0–2% mass/mass
(RTD Compensated)
Sodium Hypochlorite
Resolution
0.1%
Sodium Hypochlorite
Accuracy
±
2% of reading or
±
0.1%
Temperature Output
Range
4 mA: 32 °F (0 °C),
20 mA: 212 °F (100 °C)
Method
UV Absorbance
Calibration
Two-point calibration against standard solution
Outputs
2x 4–20mA Analog Output,
RS-485 Digital Output with Modbus protocol
Installation
1/4” OD (7 mm) Teflon tubing for measurement flow and custom tee
assembly (P/N: ST-001) with 3/4” female socket & NPT threaded ports
Cable Length
5 ft with IP67 connectors & 2 ft flying lead with IP67 adapter
Power Supply
22–26 VDC, 2W
Dimension (L
×
Dia)
6.8
×
1.44 inch (172.7
×
36.6 mm)
Weight
0.37 lbs (170 g)
Material
CPVC
Operational
Temperature
40–104 °F (4–40 °C)
Storage Temperature
20–140 °F (-7–60 °C)
Pressure
Up to 100 psi (0.7 MPa)
Enclosure Rating
IP67
Regulation
CE/RoHS
*
With Pyxis’s continuous improvement policy, these specifications are subject to change without notice.
3
Unpacking Instrument
Remove the instrument and accessories from the shipping container and inspect each item for any damage
that may have occurred during shipping. Verify that all accessory items are included. If any item is missing
or damaged, please contact Pyxis Lab Customer Service at
ST-600 Series User Manual
| +1 (866) 203-8397
3