4. Turn on the oxygen flow at the desired setting.
5. When turning off the
KnightStar 330
, turn off the
oxygen supply
first,
then turn off the
KnightStar 330
.
Figure 3-1. How to connect the oxygen adapter to the
outlet filter
Do not use supplemental oxygen at flows
above 15 L/min. Doing so could increase the
oxygen concentration inside the device and
increase the risk of fire.
To prevent oxygen from accumulating in the
device and tubing, turn on the device before
turning on the oxygen supply; shut off the
oxygen before turning off the device.
Oxygen adapter
Outlet filter
O2 port connection
Connect O2 tubing here