Instrument Specifications
TABLE 2:
Instrument specifications
Specification
Value
Weight
24.5 kg (54.0 lbs.)
Dimensions (W x D x H)
41.9 x 48.9 x 42.4 cm (16.5 x 19.3 x 16.7 inches)
Power requirements
North America:
120 VAC, 50/60Hz, 2.5 A with a NEMA 5-15 receptacle
Europe:
230 V
∿
50 Hz, 1.0 A (connectors vary by country)
Ventilation requirements
Minimum 5.09 cm (2 inches) around all sides
Usable temperature range
15 - 25°C (59-77°F), Indoor use only
Humidity
30-70% (non-condensing)
Fluidic Chip Specifications
Fluidic Chip Dimensions
TABLE 3:
Fluidic chip dimensions
Description
Value
Height
14.53 mm
Width
85.48 mm
Length
127.76 mm
Fluidic Chip Volumes
TABLE 4:
Fluidic chip volumes
Description
Value
Number of samples
Up to 8 samples per fluidic chip
Sample well volume
200
µ
L (typical)
Buffer well volume
Varies by protocol
Extraction well volume
> 50
µ
L
Fluidic Chip Storage and Handling
TABLE 5:
Storage and handling specifications
Description
Value
Temperature
15 to 32 °C
Packaging
In sealed, static dissipative bag
Handling
With disposable gloves, avoiding contact with surface