42
5.4 Start-up screen
The system enters the start-up screen after the self-test completed.
The start-up screen is the first screen that can be operated by users. It shown as below.
Figure 5-1 Start-up screen view
In this screen, users can perform the following operations:
1. Patient Category Selection:
【
Previous Patient
】
,
【
Adult
】
and
【
Pediatric
】
.
If you choose the “Previous patient”, the ventilator will continue the ventilation according to
the latest settings.
If you choose the “Adult” or “Pediatric”, the ventilator will provide ventilation according to
the default settings.
Different parameter setting and operation is confirmed by selecting different patient
categories.
Please follow the below procedures to set the patient category:
1) Touch or rotate the navigation wheel to select the needed patient category.
2) Press the navigation wheel to confirm setting.
If you select the needed patient category, the symbol of the selected patient category will
appears in the status message display area.