ProteOn XPR36 | Introduction
24
P
OWER
U
P
S
TATE
When communication is established and the instrument is ready to receive
commands, it goes into Power Up state. From this time on, the computer
monitors the current values of instrument parameters. Temperature control
begins as soon as the powering-up process is complete in order to stabilize
the instrument.
To leave the Power Up state:
•
Insert a sensor chip or shut down the instrument by clicking
Shutdown on the Instrument Control screen
See Shutdown Procedures on page 166.
I
NITIALIZING
S
TATE
Note:
Initializing takes 5 to 10 min, after which the state changes
to Ready.
When a sensor chip is inserted, the software identifies the chip and runs the
initialization protocol appropriate for the type of chip. The process may require
an additional step. For example, you may need to put glycerol in the correct
autosampler location.
When a chip initialization error occurs, a dialog box appears that lists possible
reasons for the failure. The experiment cannot continue to run until the
initialization error is resolved. You can retry initialization (using glycerol)
repeatedly before replacing the chip.
If a critical error occurs during Initializing, the state changes to Fault. If a
noncritical error occurs, a warning appears, but the state changes to Ready.
To recover from a Fault state:
•
Click Reset on the Instrument Control screen
R
EADY
S
TATE
You can start an experiment while in the Ready state. The instrument has
brought the chip to the required experiment temperature and is maintaining
temperature control (as noted in the status bar).
R
UNNING
S
TATE
Note:
You cannot change instrument settings in Running state.
The instrument is running an experiment based on the protocol you have
used.
Summary of Contents for XPR36
Page 1: ...ProteOn XPR36 Protein Interaction Array System Protein Interaction Analysis User Manual...
Page 2: ......
Page 3: ...ProteOn XPR36 Protein Interaction Array System Version 3 1 User Manual...
Page 4: ......
Page 10: ...ProteOn XPR36 System viii...
Page 166: ...ProteOn XPR36 System Analysis 148...
Page 188: ...ProteOn XPR36 Maintenance 170...
Page 194: ...ProteOn XPR36 System ProteOn System Troubleshooting 176...
Page 226: ...ProteOn XPR36 System A ProteOn XPR36 Instrument and Peripherals 208...
Page 250: ...ProteOn XPR36 System E Ordering Information 232...
Page 264: ...ProteOn XPR36 System F Security Edition Configuration Guide 246...
Page 276: ...ProteOn XPR36 System Glossary 258...
Page 287: ......