![ProteOn XPR36 User Manual Download Page 180](http://html1.mh-extra.com/html/proteon/xpr36/xpr36_user-manual_1620131180.webp)
ProteOn XPR36 | Maintenance
162
4. Place the knurled thumbscrew through the plunger opening and turn it
clockwise by hand or until it is finger tight.
5. Check buffer bottles to make sure they contain enough buffer. Add
buffer if necessary.
R
ESUMING
THE
W
IZARD
1. Click Next to continue the wizard. Choose which buffer bottle to use for
priming.
2. Verify that the syringes have been cleaned and/or installed, that the
buffer bottles contain sufficient buffer, and that the instrument is ready
for priming. Then press Next.
Priming takes about six minutes. While the priming step is running,
watch the syringes for any problems.
Caution:
Hand tighten the syringes only. The use
of tools can damage the syringe or the
instrument.