15.1.2 Calibrating the ORP Sensor
15.1.2.1
Selecting the calibration process for ORP
There are two calibration processes available for calibrating the
controller:
n
1-point (with buffer solution)
n
Data input
1.
Continuous display ➨
CAL ORP
Pot. equalisation
Calibration process
A1027
No
0.0 mV
Offset
Last calibration
CAL setup
13:26:11
11/04/2013
continue with <CAL>
1 point
Fig. 45: [ORP] calibration menu
ð
The calibration menu is displayed.
2.
Use to select the Setup menu or start calibration by
pressing
3.
[CAL Setup]: Press
ð
The menu for selecting the calibration process appears.
4.
Using the arrow keys select the required menu item
[Calibration process] and press
ð
The input window appears.
5.
Use the arrow keys to select the calibration process and
press
6.
Continue with
ð
You can now start your chosen calibration process.
15.1.2.2
1-point calibration of ORP sensor (CAL)
Correct sensor operation
–
Correct measuring and metering is only pos‐
sible if the sensor is working perfectly
–
Refer to the sensor's operating instructions
–
Remove the sensor from the in-line probe
housing and re-fit it for calibration. To do this,
refer to the operating instructions for your in-
line probe housing
Selecting the calibration process
Selecting the calibration process
Calibrating the controller diaLog DACb
84