![ProMinent DULCOTEST CDR 1 Assembly And Operating Instructions Manual Download Page 56](http://html1.mh-extra.com/html/prominent/dulcotest-cdr-1/dulcotest-cdr-1_assembly-and-operating-instructions-manual_1616849056.webp)
4.1 Production of electrolyte
Produce the ready-to-use electrolyte using the set supplied for independent production of electro‐
lyte. Unlike ready-to-use electrolyte, these individual components are not regarded as hazardous
materials. Take the following steps.
1 set to produce the electrolyte:
n
1 bottle of supporting electrolyte,
n
1 tube with powder,
n
1 label for the electrolyte produced.
A2723
Fig. 2: Production of electrolyte
1.
Take appropriate measures, such as protective gloves and goggles (1), to protect yourself,
according to the Material Safety Data Sheets for the components.
2.
Open the bottle of supporting electrolyte and the tube with powder (2).
3.
Guide the tube at an angle fully into the opening of the bottle of supporting electrolyte (3).
4.
Completely transfer the powder into the bottle of supporting electrolyte (4).
5.
Close the bottle and shake the bottle until the powder has fully dissolved, ≧ 10 seconds (5).
ð
No solids and streaks should be visible in the electrolyte.
6.
Remove (6) the label from the supporting electrolyte.
7.
Replace (7) the label with the label supplied.
ð
You can now use the electrolyte you have produced (8).
Assembly
56