Page 23
4
Storage and Transport
IMPORTANT
• Store and transport the DGMa in its original packaging.
• Protect the DGMa from the effects of chemicals, even when packed.
Environmental conditions
Storage and transport temperature: -10
°
C ... + 60
°
C
Humidity:
with flow sensor: max. 90 % relative humidity,
non-condensing
5
Assembly and Installation
5.1
Assembly
IMPORTANT
• Observe the flow direction (there are arrows on the modules).
• Install the in-line probe housing horizontally in an upright position.
• If it contains a flow module, install the in-line probe housing vertically. Failure
to do so may lead to problems with flow measurement.
• Leave a space of approx. 300 mm above and 100 mm below the modules for:
- Installing the sensors
- Setting the flow monitor
- Screwing in the calibration cup
- Taking samples
NOTE
Moistening the seals slightly first will make it easier to assemble the components
of the in-line probe housing.
Securing the mounting panel
(Please observe dimensioned drawing! See figure 5):
Drill 2 mounting holes in a smooth wall
Secure the mounting panel to the wall.
Installing the connections
Screw the ball valve onto the in-flow side
Screw a connector set onto the ball valve
Screw a connector set onto the out-flow side
Installing the sensors and
flow sensor
IMPORTANT
The first module must be the in-line probe housing module.
Sensor with PG 13.5 threaded connector
(please refer to the sensor operating manual):
Remove the upper blanking plug of a module
With pH and redox sensors, remove the transparent protective cap if there is one (do not
discard the protective caps)
Screw the reducing pipe nipple into the module
Screw in the sensor
With DULCOTEST
®
Perox sensors, screw the flow plug of an in-line probe housing module
into the module from below
Storage and Transport/Assembly and Installation
BA_DTZ_019_03_10_4spr.p65
31.10.2011, 8:41 Uhr
23