5.A. Setting Up a Run (continued)
2.
If the correct protocol is shown, press the Run/Stop button. The
SEV Set Up
screen will appear. The instrument
door will open, and the platform will extend, ready for sample loading.
!
Important.
At this point there are different paths forward, depending on whether the run data reporting options
are being used. See Section 5.B for more information. Proceed to Section 5.C for instructions on starting the run.
Setting Up a Run in the LEV Mode
1.
Verify that the
Home
screen indicates LEV and that the LEV hardware is installed. Select
Run
. If you configured
the instrument to record users, a list of user names will be shown. Select the appropriate user, and enter the
matching PIN, if appropriate. Press the Run/Stop button.
2.
Press the Run/Stop button. A
Verification
screen indicates which protocol was chosen.
8603T
A
Figure 22. The LEV Verification screen.
3.
If the correct protocol is shown, press the Run/Stop button. The
LEV Set Up
screen will appear. The instrument
door will open, and the platform will extend, ready for sample loading.
!
Important.
At this point there are different paths forward, depending on whether the run data reporting options
are being used. See Section 5.B for more information. Proceed to Section 5.C to start the run.
5.B. Setting Up Samples and Collecting Input Data
Setting Up Samples: No Run Data Reporting
Please refer to the DNA IQ™ Reference Kit for Maxwell
®
16 or DNA IQ™ Casework Pro Kit for Maxwell
®
16 Technical
Bulletin for detailed instructions regarding hardware mode, sample preparation, cartridge setup and purification
method.