For the complete protocol see Technical Manual, TM340, on the CD provided with your instrument or go to: 

www.promega.com/protocols

 and search for TM340.

CAUTION! 

The GloMax

®

-Multi+ Detection System weighs approximately 35lb (16kg) and should be handled by 

two people.  

2.

What’s in 

the box?

1.

Unpacking the 

instrument

GloMax

®

-Multi+ Detection  

System with Instinct

 Software

Setting Up Your Instrument. It’s Easy!

1.   Set the shipping box on a table, and remove the accessories and literature.
2.  Carefully remove the GloMax

®

-Multi+ Detection System from the box and place on a flat, level, solid surface in a 

dust-free location with good air circulation. Leave at least 7.5 inches (19cm) of clearance in front of the instrument to 
allow the instrument door to open without hindrance.

3.  Install the Instinct Software™ on a PC/Windows

®

 computer. 

The GloMax

®

-Multi+ Detection System with Instinct 

Software Accessories: 

A.

 Power Cord.  

B.

 AC Power Adapter.  

C.

 Microplate Adapter, 

located inside the instrument. 

D.

 Two Optical Well Crosstalk Masks, 

one for 96-well plates and one for 384-well plates. The 96-well Optical 
Crosstalk Mask will be located inside the instrument. 

E.

 USB Flash 

Drive (appearance may vary from Figure E).
Included but not shown: CD containing GloMax

®

-Multi+ Detection 

System Technical Manual (TM340) and Instinct software, one 
spare red Breakaway Tab for the plate mechanism and a small 
screwdriver.

Check to make sure all accessories are included with 
the instrument and are undamaged.

1.   Open the accessories box included with your shipment, 

and locate the instrument power cable.

2.   Connect the AC Power Adapter to the power inlet on the 

back of the Instrument. Connect the power cord to the 
AC Power Adapter.

3.   Plug the power cord into a grounded outlet.

3.

Setting  

up the 

instrument

4.

Starting up 

the instrument

1.  Turn On the GloMax

®

-Multi+ Detection System. The power switch is located 

on the back panel next to the power connection. An orange LED located left 
of the touch screen will illuminate. 

2.   The instrument will initialize and calibrate for about one minute, at which point 

the touch screen will activate and the Home screen will appear.

3.  In order to prevent damage to the instrument and loss of sample it is important 

to have the Microplate Adapter in the correct orientation. Please refer to 
Technical Manual TM340 for complete instructions.

Important:

 When using the instrument for the first time, you must press the 

“Door” button and remove the red labels attaching the Microplate Adapter to the 
Microplate Sample Tray.

Please refer to Technical Manual TM340 for detailed instructions on your 
specific application.

8785T

A

Continued on other side.

96 

We

ll

6-48 

We

ll

8007MA

Remove

8066T

A

E.

D.

C.

B.

A.

6998TB

RS-232

serial port

USB

port

Injector system

connection

Power

inlet

On/Off

switch

6998TB

©2011 Promega Corporation. All Rights Reserved. Printed in USA 12/11. 18284-PM-CR. Part# FB117

5295MA.eps