Promax OD-590 Quick Start Manual Download Page 1

OD-590

Series are powered by 1 GS/s real time and 

25 GS/s equivalent time sampling rate combined with 
3 types of acquisition modes. 25k points of deep memory
guide us into further signal details. 4 types of flexible triggers
add another signal capturing flexibility.

Signal detection

The last panel setting is internally stored in non-volatile
memory, ready to be recovered on the next power up . If you
want to transfer the setup to another 

OD-590

Series, 

you can save and recall setup files using USB flash drive.
You can always recover the default system setting.

Setup Recovery and Transfer

IEEE remote control commands include most of the panel
operations and the syntax conforms to IEEE 488.2 standard.
PC software allows you to use your oscilloscope utilizing the
larger PC screen. Three types of remote control interface
are provided: USB, RS-232C and GPIB (optional).

Remote Access

OD-590 

USB host connector transfers data quickly and

easily between USB flash drive, which guarantees almost 
unlimited amount of memory. Internal storage includes 
4 sets of reference waveform and 20 sets of general-use
memory area. 

OD-590

Series handle three types of data:

BMP image for viewing waveform shape and pasting into
documents and presentations, panel setting for saving and
restoring system setup, and waveform configuration for 
further analysis of signal information. Printout of display
image, is available through the printer connected to the USB
host port.

Data Transfer and Printout

A variety of measurement shortcuts drastically reduce repetitive
manual operations. Autoset automatically configures the 
horizontal & vertical scale and the trigger, giving an instant view
of almost any signal. 27 types of automatic measurements 
include voltage, frequency, and delay. 

OD-590

Series run and

update results of all the relevant measurements in real time. Add
and subtract math operation, with 4 types of FFT are also 
provided. Go/NoGo test function detects a user-defined 
incoming waveform shape, and can also send a signal to 
external devices in case of detection. 

Program & play

feature 

automatically runs predefined sequence and setup, boosting
productivity in routine measurements.

Measurement functions

Battery power operation option with typical 3 hours of 
running time gives a much-desired mobility. Built-in 
self-calibration and probe compensation help maintaining
maximum accuracy. You can make the oscilloscope talk your
language by choosing between a variety of languages.

Portability & Friendly user Interface

Large 5.6-in TFT Color Display

200 / 100 MHz Bandwidth

2 and 4 input channels

1 GS/s Real-Time and 25 GS/s Equivalent-Time Sampling

25k Points Record Length Maximum

27 Automatic Measurements

Three Math Functions: “+”, “-” and “FFT”

Multi-Language Support

USB Host: Flash Drive Storage and External Printer

USB Device: PC Remote Control

Battery Power Operation (Optional)

GPIB & RS-232C Interface

OD-590 Series

SPECIFICATIONS

OD-590

OD-591

OD-592

VERTICAL

Channels

2

2

4

Bandwidth

DC ~ 100 MHz (-3 dB)

DC ~ 200 MHz (-3 dB)

Rise Time

3.5 ns Approx.

1.75 ns Approx.

Sensitivity

2 mV / div ~ 5V/div (1-2-5 increments)

Accuracy

± (3% x |Readout| + 0.05 div x Volts/div + 0.8 mV)

Input Coupling

AC, DC & Ground

Input Impedance

1 M

± 2%, ~16 pF

Polarity

Normal & Invert

Maximum Input

300 V (DC + AC peak), CAT II

Waveform Signal Process

+, -, FFT

Offset Range

2 mV/div ~ 20 mV/div: ± 0.5 V

50 mV/div ~ 200 mV/div: ± 5 V

500 mV/div ~ 2V/div: ± 50 V

5 V/div: ± 300 V

Bandwidth Limit

20 MHz (-3 dB)

TRIGGER

Sources

CH1, CH2, Line, EXT

CH1, CH2, CH3, CH3, Line

Modes

Auto-Level, AUTO, NORMAL, SINGLE, TV, Edge,

Pulse Width Time-delay (2 CH Only), Event-Delay (2 CH Only)

Coupling

AC, DC, LF rej., HF rej., Noise rej.

Sensitivity

DC ~ 25 MHz: approx 0.5 div or 5 mV

25 MHz ~ 100 / 200 MHz: Approx. 1 div or 10 div

EXT TRIGGER

Range

+ 15 V

Sensitivity

DC ~ 30 MHz: ~ 50 mV

30 MHz ~ 100 / 200 MHz: ~ 100 mV

Input Impedance

1 M

~ 2%, ~ 16 pF

Maximum Input

300 V (DC + AC peak), CAT II

HORIZONTAL

Range

1 ns / div ~ 10 s/div (1-2-5 increments); ROLL: 250 ms/div ~ 10s/div

Modes

MAIN, WINDOW, WINDOW ZOOM, ROLL, SCN, X-Y

Accuracy

± 0.01%

Pre-Trigger

20 div maximum

Post-Trigger

1000 div

X-Y MODE

X-Axis Input

Channel 1 (X-Axis Input); Channel 2 (Y-Axis Input)

Phase Shift

±3º at 100 kHz

SIGNAL ACQUISITION

Real-Time Sample Rate

1 GS/s maximum

Equivalent Sample Rate

25 GS/s maximum

Vertical Resolution

8 bits

Record Length

25 K Dots maximum

Acquisition Mode

Normal, Peak Detect, Average

Peak Detection

10 ns

Average

2, 4, 8, 16, 32, 64, 128, 256

CURSORS AND MEASUREMENT

Voltage Measurement

V

pp

, V

amp

, V

avg

, V

rms

, V

hi

, V

lo

, V

max

, V

min

, Rise Preshoot / Overshoot, Fall Preshoot / Overshoot

Time Measurement

Freq, Period, Rise time, Fall Time, Positive Width, Negative Width, Duty Cycle

Delay Measurement

Eight different delay measurement

Cursors Measurement

Voltage difference between cursors (

V) Time Difference between cursors (

T)

Auto Counter

Resolution: 6 digits; Accuracy: ± 2%

Signal Source: All Available trigger source except the Video Trigger mode

CONTROL PANEL FUNCTION

Autoset

Adjust Vertical VOLT/DIV, Horizontal TIME/DIV, and Trigger level automatically

Save Setup

Up to 20 sets of measurement conditions

Save Waveform

24 sets of waveform

DISPLAY

Color TFT LCD Type

5.6 inch

Display Resolution

234 (Vertically) x 320 (Horizontally) Dots

Display Reticule

8x10 divisions; 8x12 divisions (menu off)

Display Brightness

Adjustable

INTERFACE

Go/NoGo Output

5 V Maximum / 10 mA TTL Open Collector Output

RS-232 Interface

DB 9-pin male DTE RS-232 interface

GPIG Interface (Option)

Fully programmable with IEEE 488.2 compliance

USB

USB Host/Device 2.0 full speed supported

POWER SOURCE

Line Voltage Range

AC 100 ~ 240 V, 48 ~ 63 Hz, Auto selection

Battery Power (Option)

Battery: 11.1 V Li-Ion battery pack

Charge Time: 8 hours (Power ON); Operating Time: 3 hours, depending on operating condition

MISCELLANEOUS

Multi-Language Menu & Online Help

Available

Time Clock

Time and Date, Provide the Date/Time for the saved data

DIMENSIONS & WEIGHT

243 D. x 142 H. x 310 W. (mm); approx. 4.3 kg

Bandwidth

Channels

Display

Sample Rate

Record length

Trigger

Valued features

Interfaces

OSD

Power Source

100 MHz

2

5.6” TFT Color LCD

1 GS/s

25k pts

Pulse Width, TV Line,

Event Delay, Time Delay

Multi-Language, FFT, Go/NoGo,

Auto Setup, 27 Auto-Measures

2xUSB (Host/Device), RS-232C,

Go/NoGo Output, GPIB (optional)

Multi-Language Menu

Multi-Language Online Help

AC 100 ~ 240 V

Optional Li-Ion battery pack

200 MHz

2

200 MHz

4

outstanding specifications

OD-590

OD-591

OD-592

OD-592

200 

MHz

channels

PROMAX ELECTRONICA, S.A.

Francesc Moragas, 71  *  08907 L'HOSPITALET *  SPAIN
Tel: (+34) 93 260 20 02  *  Fax: (+34) 93 338 11 26  *  e-mail: [email protected]  *  www.promax.es

S

pecifications subject to change without notice.  0 IP4584

OD-591

200 

MHz

channels

OD-590

100 

MHz

channels

200

MHz