Page 56
14 Built-in battery and charging
14.1 Battery
The built-in Li-Ion-Mn battery has a very large capacity which is monitored
continuously. When the state of charge is running low the battery symbol is
shown on the display and the battery has to be charged.
In order to prevent damage to the battery, the instrument
switches off automatically before the battery is completely
drained. The ZRS 6060 announces this by displaying the
symbol
and a message.
In order to preserve the battery,
the „Auto off time“ can be
changed, as described in chapter 13.6.4
“Power” on page
14.2 Battery status indication on the display
Battery is empty. The instrument switches off automatically.
Battery is soon empty. The instrument should be charged.
approx. 50%
approx. 80%
100% battery is fully charged
Instrument is connected to the battery charger -> the charging status is indicated
on the battery charger.
Charging cycle
The charge indicator light on the charger shows the charging status of the battery in
the charging cycle:
•
red: the unit is fast charging. The charge current is maximum and the charger
is in continuous mode.
•
orange:
the charging cycle “final charge” is active. The battery is now charged
at approx. 80 %. The battery charger is in continuous mode.
•
green: the battery is fully charged and the charger is in trickle charge mode.
14.3 Charging