16 Status and error messages
16.1 Status message
– “Battery voltage is critical”
In order to prevent damage to the battery, the
instrument switches off automatically before
the battery is completely drained. The ZRP
6030+ announces this by displaying the
symbol and a status message.
Charge the battery as described in chapter
14.3
“Charging” on page 43.
16.2 Status message
– “Please calibrate R”
Press the warning triangle to get a more
detailed explanation.
To solve this, carry out the calibration as
described in chapter 7
“Calibrate” on page
16. If you are sure that it is not necessary to
carry out a calibration, you can press
“Acknowledge all” and the message will
disappear.
To change the calibration interval refer to
chapter 13.6.8
“Calibrate” on page 40.
16.3 Error message
– “Calibration is not OK”
Calibration results with deviations greater
than 15 % are indicated by “Calibration is not
OK”. This can be caused by a soiled or
damaged standard or a soiled optic window.
It is suggested to reject such a calibration
result by pressing
before determining
the reason for the deviation. Clean the
standard as described in chapter 7.2
“Cleaning of the calibration standard” on
page 17 and recalibrate. When repeatedly
experiencing a deviation greater than 15%,
service or factory calibration might be
required. Please contact Proceq or your