Meter Testing
Tests performed with a pickup accessory (Optical or Metercam)
Five different tests can be done with optical pickups:
Quick
,
Full
,
Custom, Four-
Quadrant
(WH)
and
Four-Quadrant (VARH)
Test.
Quick Test
,
Full Test, Four-Quadrant (WH)
and
Four-Quadrant (VARH) Test
are made
up of a predetermined sequence of consecutive steps and tests, all of which are done
in one operation. The sequence is programmable and can be modified through the
Settings
tab. By default, the sequence includes one test point with each of the loads:
HL, PF and LL.
In
Quick Test
,
the three phases (ABC) are tested simultaneously in one step.
In
Full Test
,
a
Quick Test
is performed followed by each phase separately (A, B, and
C). In the
Settings
tab, there is an option to remove the PF and/or LL for each single
element test.
In
Four-Quadrant (WH) Test
, the equivalent of a Quick Test is done at various phase
lags between Voltage and Current. This is meant to test the operation of the meter
in all four quadrants of the power vector diagram. For this test, the device expects
the meter to pulse on WH units.
In
Four-Quadrant (VARH)Test
, the equivalent of a Quick Test is done at various phase
lags between Voltage and Current. This is meant to test the operation of the meter
in all four quadrants of the power vector diagram. For this test, the device expects
the meter to pulse on VARH units.
Please note that with single-phase meters, only the
Quick Test,
Four-Quadrant (WH)
and
Four-Quadrant (VARH) test
will be activated as there is no
Full Test
for single-
phase meters.
Custom Test
consists in choosing one of the three loads to apply to the meter: HL, PF
or LL. The disk revolutions or pulses are counted automatically by the pickup. No
critical timing is required to start a test. The test ends automatically.
When testing a three-phase meter, you can also choose to apply the load on all
phases (ABC) at the same time or on each phase individually (A, B or C). The phase
selection is done by selecting the appropriate phase (A, B, C, or A+B+C) in the
Select
Active Phase
drop down window.
Tests performed without a pickup accessory (Optical or Metercam)
Manual Test
is used when the operator counts the disk revolutions manually. A
magnet is used to start and end the test manually by applying it to the side of the
socket (see figure 2.2). The operator can set his own parameters for conducting a
test, such as the number of revolutions and the load to be applied. The load can be
switched dynamically during the test.
Summary of Contents for MT-1/WT3
Page 1: ...MT 1 WT3 USER GUIDE HIGH ACCURACY THREE PHASE WIRELESS WATTHOUR METER TESTER...
Page 9: ...Probewell Lab Inc Page 8...
Page 12: ...MT 1 WT3 User Guide Page 11 Side View of the Socket Fig 2 2 Side View of the Socket...
Page 21: ...Probewell Lab Inc Page 20...
Page 43: ...Probewell Lab Inc Page 42...
Page 47: ...Probewell Lab Inc Page 46...
Page 49: ...Probewell Lab Inc Page 48 Form 1S...
Page 50: ...MT 1 WT3 User Guide Page 49 Form 2S...
Page 71: ...Probewell Lab Inc Page 70 Form 12S 25S 3N N Network...
Page 72: ...MT 1 WT3 User Guide Page 71 Form 12S 25S 3 25S Delta...
Page 73: ...Probewell Lab Inc Page 72 Form 14S 4Y s...
Page 74: ...MT 1 WT3 User Guide Page 73 Form 15S 4...
Page 75: ...Probewell Lab Inc Page 74 Form 16S 4Y...
Page 77: ...Probewell Lab Inc Page 76...
Page 83: ...Probewell Lab Inc Page 82...
Page 85: ...Probewell Lab Inc Page 84 Notes...
Page 86: ...www probewell com Probewell Lab Inc Printed in Canada Edition 2 2...