Operational Forces
Handset
4N
Emergency Button
4N
Hook locking mechanisms on lift tape
2.5N
Spring clips on carry bar
8N
Manually traversing fully loaded hoist (SWL)
97N
Manually traversing unloaded hoist (No weight)
6N
Table 9
7.3
Expected Product Lifetime
Ten years depending usage and compliance to maintenance, servicing and LOLER inspections.
Serviceable parts within this period are batteries and the lift tape. Batteries should have an expected service life of >400
discharge cycles, dependant on the charging routine. The lift tape should have an expected service life of 2 years if used
correctly but visual inspection should be carried out before use.
7.4
Standards Applied
The standards that have been applied to the device are as follows:
•
EN 10535:2006 Hoists for the transfer of disabled persons. Requirement and test methods
•
EN 60601-1-1:2006 +A12:2014 Medical electrical equipment. General requirements for basic safety and essential
performance
•
EN 60601-1-2:2015 Medical electrical equipment. General requirements for basic safety and essential performance.
Collateral Standard. Electromagnetic disturbances. Requirements and tests
•
EN 60601-1-6:2010 Medical electrical equipment. General requirements for basic safety and essential performance
•
EN 60601-1-11:2015 Medical electrical equipment. General requirements for basic safety and essential performance
8.0
Environmental - Storage and Operating Conditions
The Hoist is intended for internal use within normal environmental conditions.
It is not intended to be used in environments where there are rapid changes in the environmental temperature and
humidity during intended use.
•
Lint
–
Due to the nature of the CP Hoist being installed closely to the ceiling, very little lint would be likely to gain
access into the
hoist’s
workings. The hoist is recommended as per Service Guide to be wiped cleaned during every
hoist inspection.
•
Dust
–
Due to the nature of the CP Hoist being installed closely to the ceiling, very little dust would be likely to gain
access into the
hoist’s
workings.
•
Light
–
The User controls have been designed to be easily recognisable and the use of bright colours will help the
user through all ranges of lighting. The Specification of the hoist dictates that normal use would occur during
ambient luminance 50
–
500 lux. Additional as the hoist is designed for indoor use only, if required the user may wish
to switch on room lighting.