background image

 

 

 

 

 

999548 – Revision B 

 

P a g e

 | 

6 of 20

 

 

 

Ensure that a clear space is maintained around the FSG440. Before performing a transfer, check for and move all 

obstacles out of the way. 

 

 

Your Product is for aiding and transferring a patient. Do not use it, or allow it to be used, for any other purpose. 

 

 

In areas where children are prone to be present be vigilant when carrying out a transfer.  

 

 

To reduce the risk of unintended use, when the FSG and Hoist is not in use, remove the sling(s) from the product to 

prevent entrapment or strangulation should the device be tampered with. 

 

 

Between FSG440, Hoist, Spreader Bar and Body-Support Unit, the lowest maximum load shall always be used. 

 

You may need to seek specialist advice on how to assist some people with specific moving and handling needs. 

Sources of advice include, but is not limited to, professional bodies and organisations, occupational therapist, 

physiotherapists, manual handling advisers and ergonomist with experience in health and social care. 

1.7

 

Warning Notes 

Your Free-Standing Gantry has been manufactured and tested to exceed BS EN 10535:2006. This does not mean that 

it can be used without care. ALL OPERATORS should have read the operating instructions and appreciate this 

warning section. 

1.

 

FSG440 are less stable on sloping surfaces. Mackworth recommend that the install of the FSG440 is on a flat and level 

surface. 

2.

 

ALL FSG440 are dangerous to the person being transferred when used without training. 

 

If, during the use of this device or as a result of its use a serious incident has occurred, please report it to the 

manufacturer and to your national authority 

1.8

 

Operating Environment

 

The FSG440, Free-Standing Gantry is suitable for use within the professional health care facility 

environment as well as the following listed:  

 

 

Home environment 

 

Hospital and/or treatment centre  

 

Residential care home  

 

Education institute 

 

The FSG440 is not suitable for any special environments.  

This product is not intended to be used in environments where there are rapid changes in the 

environmental temperature and humidity during intended use. 

1.9

 

Labeling  

The serial number label shown in figure 1 is located on both support posts. The 

information which is present is: 

 

Manufacturer address  

 

Manufacturing Date 

 

Reference number Serial number  

 

Safe working load  

 

Figure 1

 

Summary of Contents for mackworth FSG440

Page 1: ...Doc No 999548 Revision of document B Revision Date 09 12 2021 FSG440 Portable Gantry User Manual...

Page 2: ...s Key Parts 7 Final Inspection 8 Technical Specification 9 4 1 FSG Gantry Dimensions 9 4 2 Specifications 9 4 3 Expected Product Lifetime 10 4 4 Standards Applied 10 Environmental Storage and Operatin...

Page 3: ...with the documents included with the FSG440 Contents of this manual are subject to change without prior written notice Do not attempt to use this equipment without first understanding the contents of...

Page 4: ...l represents Consult instructions before use Caution see instructions for use Manufacturer Safe Working Load represents the maximum load rated for safe operation Packaging indicator This way up Date o...

Page 5: ...flat and levelled surface prior to use The FSG440 must be installed only by persons authorized by Mackworth or who have the rights to install and commission the Gantry safe for use Under no circumstan...

Page 6: ...tested to exceed BS EN 10535 2006 This does not mean that it can be used without care ALL OPERATORS should have read the operating instructions and appreciate this warning section 1 FSG440 are less s...

Page 7: ...ystem The image below shows the contents of the free standing gantry If you have not received all the components contact your Mackworth dealer immediately contact details are provided on the last page...

Page 8: ...ist accessory and sling user guides for specific details regarding their inspection Prior to mounting the portable lift onto the trolley of the FSG440 please read the owner s manual of the portable li...

Page 9: ...8 7 Lowest Height 2141 mm 84 3 Maximum length 3387mm 133 3 Minimum Length 2445mm 96 3 Internal maximum clearance 3052mm 120 2 Internal minimum clearance 2102mm 82 8 Operational Forces Latch 3N Moving...

Page 10: ...up to 90 35 C to 70 C 95 F to 158 F non condensing at a water vapour pressure up to 50hPa Disposal When the FSG440 has completed its life cycle and can no longer perform to its intended use safely the...

Page 11: ...arried out on a Mackworth FSG440 while transferring a person to reduce the risk of injury Service must be completed by a Mackworth authorised Service Engineer Do not attempt to service the product you...

Page 12: ...ised dealer immediately Certification An authorised competent service company or person will issue a test certificate after satisfactory completion of the LOLER inspection The certificate will be vali...

Page 13: ...s of the device should be for 5 minutes to prevent any virucidal infections without a degradation to the functionality of the device 5 minutes is a recommended contact time The device can withstand a...

Page 14: ...MSDS Manufacturers Safety Data Sheet For a full review of the data please follow the link below http www nhsggc org uk media 236215 msds actichlor plus pdf Warranty This guarantee does not affect or...

Page 15: ...999548 Revision B P a g e 15 of 20 10 0 Service Record History Complete this section after each service repair inspection and or maintenance...

Page 16: ...999548 Revision B P a g e 16 of 20...

Page 17: ...999548 Revision B P a g e 17 of 20...

Page 18: ...999548 Revision B P a g e 18 of 20...

Page 19: ...999548 Revision B P a g e 19 of 20 User notes...

Page 20: ...y errors or omissions Mackworth operates a policy of continuous improvement Specifications and other data are subject to change without notice Dealer service contact details Manufactures Contact detai...

Reviews: