ten minutes. Although the thermo-protection switch will protect the camera, you are
nevertheless advised to power down and correct the operating conditions that caused
the thermal-overload to occur.
Gradual Deterioration of Cooling Capability
While unlikely with the PI-MA3 camera (guaranteed permanent vacuum for the life of
the camera), if you see a gradual deterioration of the cooling capability, it might due to
damaged camera vacuum. This can affect temperature performance such that it may be
impossible to achieve temperature lock at the lowest temperatures. In the kind of
applications for which cooled CCD cameras are so well suited, it is highly desirable to
maintain the system’s lowest temperature performance because lower temperatures result
in lower thermal noise and better the signal-to-noise ratio. Contact the factory to make
arrangements for returning the camera to the support facility.
Data Loss or Serial Violation
You may experience either or both of these conditions if the host computer has been set
up with Power Saving features enabled. This is particularly true for power saving with
regard to the hard drive. Make sure that Power Saving features are disabled while you are
running WinView/32 or WinSpec/32.
Error Occurs at Computer Powerup
If an error occurs at boot up, either the Interface is not installed properly or there is an
address or interrupt conflict. Turn off the computer, try a new address or interrupt and
reinstall the card. Be sure the Interface is firmly mounted in the slot.
Since interrupts and DMA channels cannot be shared, make sure no other boards in your
computer use this interrupt or these DMA channels.
one and reports the address and interrupt assignments for that one as well. In a few
Ethernet Network is not accessible
When the Princeton Instruments software is installed, all Intel Pro/1000 interface card
drivers found on the host computer are updated with the Intel Pro/1000 Grabber Adapter
(Vision High-Performance IP Device) driver provided by Pleora Technologies, Inc. If
this computer is connected to an Ethernet network via an Intel Pro/1000 card that does
not use the Pleora driver, the network connection will be broken.
To restore the driver for the Ethernet card that is used to connect to the network:
1.
Locate the EbDriverTool.exe file.
Download the
EbDriverTool.exe
file to your computer from
ftp://ftp.princetoninstruments.com/public/software/official/winx32/v2525/patches
Or open the default Pleora directory. Typically it is located at
C:\Program
Files\Common Files\Pleora Technologies Inc.
The
EbDriverTool.exe
file may
be in a subdirectory.
2.
Run the file.
3.
When this program executes, select the appropriate Ethernet card and under the
Action
category, choose "
Install Manufacturer Driver
” from the pulldown menu.
Caution
Summary of Contents for PI-MAX 3 System
Page 1: ...4411 0129 Version 1 A September 8 2010 4411 0069 ...
Page 32: ...32 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 52: ...52 PI MAX 3 System Manual Version 1 A Figure 20 Safe Mode and Fast Mode Operation ...
Page 72: ...72 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 90: ...90 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 100: ...100 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 110: ...110 PI MAX 3 System Manual Version 1 A Figure 68 Outline Drawing PI MAX3 with F mount Adapter ...
Page 114: ...114 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 120: ...120 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Page 140: ...140 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...