Chapter 4
First Light
29
4.
On the
Gating
tab page, select
Repetitive
and
then click on the
Setup…
button.
5.
On the
Repetitive Gating Setup
dialog, enter
the following settings:
Gate Width:
50 ms
Gate Delay:
10 ns
Gates per Exposure:
1
Repeat Width/Delay:
1.
6.
Click on
OK
.
Acquisition menu
1.
From the Acquisition menu, select Experiment
Setup.
2.
On the
Main
tab page, make the following
settings.
CCD Readout:
Use Full Chip
Intensifier
Gain:
1
Gate Mode:
selected
3.
On the
ADC
tab page, select the appropriate
Rate
(16 MHz for Kodak 1024x1024, 1 MHz for
CCD30-11).