![Primes MicroSpotMonitor Plus HighBrilliance Original Instructions Manual Download Page 51](http://html1.mh-extra.com/html/primes/microspotmonitor-plus-highbrilliance/microspotmonitor-plus-highbrilliance_original-instructions-manual_1601728051.webp)
51
Revision 00 EN - 02/2022
MSM+ HB
35,0(6
9.4
General procedure for measuring
This chapter contains general information about the LaserDiagnosticsSoftware LDS. Read this general infor-
mation before turning to the following chapters on the different measuring tasks.
9.4.1
Open the „Device control“
menu
After connecting the device, the
Device control
menu must be opened:
1. Click on the
Devices
tab.
2. Select the device and click on the de-
vice function
CCD
below the device
name.
The
Device control
menu with the
Adjustment
and
Measuring modes
sections opens.
Device function
9.4.2
Move camera housing to desired z-position
The
Adjustment
area of the
Device control
menu offers various options for moving the camera housing.
Option
Explanation
z-Position in mm
X
Enter the desired z-position in the input field or use the slider.
Move to axis position
X
Click this button to move the camera housing to the z position spe-
cified in the
z-Position in mm
field.
Move to axis center position
X
Click this button to move the camera housing to the center position
of the (unlocked) z-axis traversing range.
Note that this does not have to be the maximum traversing range of
your device. If a locked area of the z-axis was stored in the EEPROM
during a previous use of the device, then this locked area is active. The
camera housing then moves to the center position of the unblocked
area.
Tab. 9.2:
Options in the
Adjustment
area of the
Device control
menu
Summary of Contents for MicroSpotMonitor Plus HighBrilliance
Page 2: ......
Page 3: ...3 Revision 00 EN 02 2022 MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Page 7: ...7 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 95: ...95 Revision 00 EN 02 2022 MSM HB 35 0 6 13 Declaration of conformity...
Page 96: ...96 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 98: ...98 Revision 00 EN 02 2022 MSM HB 35 0 6 15 Dimensions 15 1 Dimensions of the MSM HB10 A...
Page 99: ...99 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 continued View A...
Page 100: ...100 Revision 00 EN 02 2022 MSM HB 35 0 6 15 2 Dimensions of the MSM HB10 with fibre bridge A...
Page 103: ...103 Revision 00 EN 02 2022 MSM HB 35 0 6 15 3 Dimensions of the MSM HB20 at zmax C...
Page 104: ...104 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 continued View C...