MSM+ HB
35,0(6
8.4
Safety Circuit
Connect this port to your laser system (Safety PLC or Emergency off).
The Safety Interlock protects the MSM+ HB from damage by switching off the laser in the following cases:
• The voltage supply at the MSM+ HB is not connected or interrupted
• There is an overtemperature at the absorber
• A referencing procedure is triggered during the measurement
The interlock connection provides two redundant safety circuits.
NOTICE
Danger of damage due to overheating
If the safety circuit is not connected, the device can be damaged due to overheating.
X
Only operate the device with a correctly connected safety circuit
X
Check that the laser is switched off properly by the external safety circuit in the event of a
fault.
M8 connector 8-pin (Pin: view connector on device; Color: color of cable core)
1
8
2
7
3
6
4
5
Pin
Color
Function
1
White
COM 1: Connected with pin 2 when not ready for operation
Connected with pin 3 when ready for operation
2
Brown
NoClosed 1: Connected with pin 1 when not ready for operation
3
Green
NoOpen 1: Connected with pin 1 when ready for operation
4
Yellow
COM 2: Connected with pin 5 when not ready for operation
Connected with pin 6 when ready for operation
5
Grey
NoClosed 2: Connected with pin 4 when not ready for operation
6
Pink
NoOpen 2: Connected with pin 4 when ready for operation
7
Blue
+ 24 V, supply from external
8
Red
GND to external voltage (Pin 7)
Matching ready-made cable: Binder 79-3808-52-08
Tab. 8.2:
Pin assignment of the safety interlock connector
If the safety interlock is triggered, pin 1 and pin 2 (pin 4 and pin 5) are connected. If the values correspond to
the operating conditions, pin 1 and pin 3 (pin 4 and pin 6) are connected.
Pin 7 and 8 must be connected. Otherwise the interlock circuit is not active.
Summary of Contents for MicroSpotMonitor Plus HighBrilliance
Page 2: ......
Page 3: ...3 Revision 00 EN 02 2022 MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Page 7: ...7 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 95: ...95 Revision 00 EN 02 2022 MSM HB 35 0 6 13 Declaration of conformity...
Page 96: ...96 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 98: ...98 Revision 00 EN 02 2022 MSM HB 35 0 6 15 Dimensions 15 1 Dimensions of the MSM HB10 A...
Page 99: ...99 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 continued View A...
Page 100: ...100 Revision 00 EN 02 2022 MSM HB 35 0 6 15 2 Dimensions of the MSM HB10 with fibre bridge A...
Page 103: ...103 Revision 00 EN 02 2022 MSM HB 35 0 6 15 3 Dimensions of the MSM HB20 at zmax C...
Page 104: ...104 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 continued View C...