1. Connect water in to PLM.
2. Connect water out at PLM to water in at MSM+ HB, using the supplied hose.
3. Connect water out at MSM+ HB to cooling.
In the case of an MSM+ HB10, the temperature sensor embedded in a T-adapter must be installed be-
tween the MSM+ output and the chiller.
The lengths of the hose has an influence on the time constants of the measuring device. Only use
the hose supplied!
16.2.2
Electrical connections
Connect USB
Connect the USB port of the PLM to the PC or network using the supplied USB-cable.
Connect temperature sensor
• MSM+ HB10:
Install the temperature sensor, which is integrated within a T-adapter, between the MSM+ output and the
chiller
• MSM+ HB20:
Connect the permanently mounted temperature sensor cable (black cable on the water return of the
MSM+ HB20) to the PowerLossMonitor PLM
Connect interlock
Connect the interlock of the PowerLossMonitor PLM to the Safety-Circuit-PLM-In of the MSM+ HB. For this
purpose, use the enclosed 8-pin cable which is pre-assembled on both sides.
16.2.3
Transport or storage of the PowerLossMonitor PLM
The cooling circuit of the PLM must be drained completely before transport or storage. Temperatures close
to or below freezing point can lead to damages due to the formation of ice, if the cooling circuit is not com-
pletely drained.
NOTICE
Damage of the turbine
The turbine is not designated for high rotational speed.
X
Do not use compressed air to drain the cooling circuit of PLM.
Unlike the MSM+ HB, the cooling circuit of the PLM is not allowed to be drained with compressed air.
1. Remove cooling circuit hoses.
Cooling circuit connectors
2. Drain the cooling circuit completely.
3. Seal the connectors of the cooling circuit us-
ing the enclosed plugs.
Summary of Contents for MicroSpotMonitor Plus HighBrilliance
Page 2: ......
Page 3: ...3 Revision 00 EN 02 2022 MSM HB 35 0 6 IMPORTANT READ CAREFULLY BEFORE USE KEEP FOR FUTURE USE...
Page 7: ...7 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 95: ...95 Revision 00 EN 02 2022 MSM HB 35 0 6 13 Declaration of conformity...
Page 96: ...96 Revision 00 EN 02 2022 MSM HB 35 0 6...
Page 98: ...98 Revision 00 EN 02 2022 MSM HB 35 0 6 15 Dimensions 15 1 Dimensions of the MSM HB10 A...
Page 99: ...99 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB10 continued View A...
Page 100: ...100 Revision 00 EN 02 2022 MSM HB 35 0 6 15 2 Dimensions of the MSM HB10 with fibre bridge A...
Page 103: ...103 Revision 00 EN 02 2022 MSM HB 35 0 6 15 3 Dimensions of the MSM HB20 at zmax C...
Page 104: ...104 Revision 00 EN 02 2022 MSM HB 35 0 6 Dimensions of the MSM HB20 continued View C...