HP-MSM-HB
35,0(6
13.4
Prepare measurement
The following check lists should help you to realize the most important conditions for a measurement and to
carry out all necessary settings of the LaserDiagnosticsSoftware LDS.
13.4.1
Check list measurement settings
The device is stable and fixed.
The movement range (z-axis) of the measuring device is free of obstacles.
Tab. 13.2: Check list safety precautions
13.4.2
Check list measurement settings
LDS Menu Path
Action
Measurement > Environment
Enter the focal length
Measurement > Sensor param-
eters
Preset 64 pixel for the resolution x
Preset 64 pixel for the resolution y
Measurement > CCD Settings
Select the trigger mode Cw/Quasi-cw Measurement
Select the CCD mode Measuring data
Select the correct wave length
Measurement > Single
Activate the check box Optim.
Measurement > Caustic
Activate the mode Automatic
Activate the check box Optim.
Measurement > Option
Preset the fillfactor Max: 0.7 Min: 0.5 Ref: 0.6
Analyse Settings: Activate the check box Enable adjust nullevel
Tab. 13.3: Check list measurement settings