![Primes FocusMonitor FM+ Operating Manual Download Page 31](http://html1.mh-extra.com/html/primes/focusmonitor-fm/focusmonitor-fm_operating-manual_1601726031.webp)
FocusMonitor FM+
35,0(6
31
Revision 03/2019 EN
11.4.3
Configuring advanced settings (
Device control > Advanced
)
1. Click on the
Advanced
tab.
2. Enter the number of pixels in the
x/y-direction in order to configure the
resolution on the CCD sensor.
3. Enter the number of planes.
Specifying 21 planes already gener-
ates a valid measurement.
4. Enter in the rotation speed of the
measuring tip 1875, 3750, 7500 min
-1
.
Locked area
X
Define a locked area when there are
obstacles in the measuring area.
• More detailed information on locked
areas can be found in chapter 11.4.4
Move axes
X
To align the device (see Fig. 7.6 on
page 18) or to initiate a measure-
ment at this position, enter 60 mm for
the z-position.
• The vertical carriermoves to the speci-
fied position.
• More detailed information on travers-
ing the axes can be found in chapter
Measuring tip
You will find the entry values on the label
attached to the packaging of the measur-
ing tip used.
5. Choose the type of measuring tip
used.
6. Enter the serial number S/N of the
measuring tip.
7. Enter the sensitivity of the measuring
tip used in cts/(MW/cm²).
8. Enter the pinhole diameter of the
measuring tip used.
Rotated measuring tip
X
Click on one of the
< >
buttons if
you are using the FocusMonitor FM+
mounted upside down with a rotated
measuring tip.
The top of the measuring tip always
has to be facing the incident laser
beam.
• More detailed information on a ro-
tated measuring tip can be found in