Absorber temperature in °C
Min. energy input in J
Max. energy input in J
20
200
4 000
25
200
4 000
30
200
4 000
35
250
4 000
40
300
4 000
45
400
4 000
50
500
4 000
55
650
4 000
60
800
4 000
65
1 000
3 700
70
1 500
3 200
Tab. 12.1: Absorber temperature with the recommended minimum and permissible maximum
energy input (ambient temperature 20 °C)
12.3
Waiting time until the next measurement in a measurement series
Please note that at high measuring frequencies the measuring accuracy may be limited.
For measurements within the specified accuracy we recommend a maximum repetition