![PreXion PreXion3D Excelsior Service Manual Download Page 17](http://html1.mh-extra.com/html/prexion/prexion3d-excelsior/prexion3d-excelsior_service-manual_1599689017.webp)
XTR-PX2-C0006-IN004
1-6
1.5 Power requirements (for Scanner)
(1) Consumption current of the device
Make sure that the device that supplies power to the PreXion3D Excelsior has an apparent resistance value of
0.1 Ω or less.
(2) Power cable
Use a power cable to connect the distribution board to the PreXion3D Excelsior that meets the following
conditions:
・
Thickness: AWG14 or more
・
A cable that conforms with the local electricity safety standards (for America, that is the NEC)
* Be sure to ground the device.
(3) Installing the additional breaker
The PreXion3D Excelsior is a permanently installed device and is supplied power by the distribution board in
the building.
To ensure safety, connect the electric cable that connects the distribution board to the device to a dedicated
circuit breaker in between. Prepare a dedicated circuit breaker that is 15 A.
Before a service engineer performs maintenance, make sure to turn off this circuit breaker.
< Arrangement example >
The PreXion3D Excelsior unit is equipped with the following current interruption characteristics:
・
Standard current rating: 15 A or more
・
Rated interruption capacity: 1000 A / AC 240 V or more
Breaker
Breaker