XTR-PX2-C0006-IN004 X-ray door interlock
1
X-ray room Door Interlock kit
1. contents of the kit
There are three parts.
2. how to install the kit to the Excelsior scanner unit
2-1 15m cable assy
At the foot of the Pillar, there is a short pin for door interlock connection.
Remove the short pin connector, and connect the PA connector of 15m cable assy here.
2-2 magnet
Fasten the magnet on the sliding door ( moving door ).
2-3 proximity sensor assy
Fasten the proximity sensor assy just outside of door
proximity
sensor assy
magnet
PA connector
short pin connector
15m cable assy
proximity sensor
model: GLS-1
manufacturer: OMRON