The readjustment of the calibrator must be performed in conditions of temperature
and humidity reference.
The minimum warm-up is two hours.
The battery charger must be disconnected.
The Standards presented to the calibrator during the readjustment should have an
accuracy at least 3 times better than the accuracies of the calibrator provided in this
manual.