Page 50
Options for
OUT
adjustment are:
V
mA
There is no order or interdependence between the adjustments.
4.1. Input Adjustment
Select the corresponding mnemonic and apply the signals presented in the
tables below.
When calibrating inputs, the display shows on the 2
nd
line the value measured
by the calibrator and on the 1
st
line the same value is expressed as a percentage.
Note that the applied signals just need to be close to the values shown in the
table.
Once the signal has been applied, store the values of the 1
st
and 2
nd
adjustment points by pressing keys 1 (1
st
point) and 2 (2
nd
point).
V Input
1
st
point
2
nd
point
Single range
0.00000 V
11.00000 V
mA Input
1
st
point
2
nd
point
Single range
0.0000 mA
20.0000 mA
presys