Page 49
4 - Adjustment
Warning: Enter the following options only after understanding them
completely. Otherwise, it may be necessary to return the instrument to the
factory for readjustment!
Select
ADJ
option from the main menu and press ENTER. Enter the
PASSWORD
9875 to access the calibration menu.
The password works as a protection to adjustment ranges. After entering the
password, the menu displays the options:
IN
OUT
DATE
Then, choose if the calibration will be performed over an input range (
IN
) or an
output range (
OUT
).
DATE
is an option which allows you to record the date in which
the calibration is performed and once it has been filled in, it will be displayed every
time the instrument is turned on. The date can only be updated after an adjustment
operation.
Options for
IN
adjustment are:
V
mA
PROBE
PRESS
presys