Page 47
3.10. MEM Command
The calibrator offers several programming and special functions that can be
often used. In this case, storing these configurations in the calibrator saves time. One
can have up to eight sequences stored in memory.
After making a specific operation in the calibrator through the keyboard, return
to the menu
CONF
. Select
MEM
and press ENTER. The display will show:
WRITE
RECALL
CLEAR ALL
Select
WRITE
and press ENTER. The display will show:
1
2
3
4
5
6
7
8
The numbers shown above represent eight memory positions. Select any of
them and press ENTER. The calibrator preset configuration is stored in the chosen
memory. In order to call it back, even though the calibrator has been turned off and
on, select
RECALL
(ENTER) and the memory number that stored the previous
configuration. Press ENTER.
Any new configuration can be written over an already used memory location.
When you want to clear all eight memory locations, select
CLEAR ALL
and
press ENTER.