Page 39
is pressed, an ascending cycle starts, and by pressing
, a descending cycle begins, only
once. Pressing
and
, the cycles are repeated continuously.
3.9. Special Functions
Selecting
FN
the display shows:
INPUT
OUTPUT
Through these options, you may select special functions related to the
INPUT
or the
OUTPUT
.
INPUT has the
SCALE
,
CAL
and
NO
options.
The input type must be previously configured; otherwise the
SELECT INPUT
FIRST
message will be displayed. In this case, press
C/CE
to go back to the main
menu and select the input type.
3.9.1. SCALE Function (IN)
It establishes a linear relationship between the calibrator input signal and what
is shown on the display, according to the graphic below.
The scaled display indication (#) may represent any engineering unit, such as
m/s, m
3
/s, % etc.