Page 37
3.8.3. SPEED Programming
This program lets you modify the conversion rate of the input pressure. By
selecting
YES,
the conversion rate becomes twice as fast as without this option.
3.8.4. STEP Programming
The
STEP
programming makes the calibrator output vary in pre-defined
steps. It is useful in calibrations where certain scale points are verified, e.g.,
0% - 25% - 50% - 75% - 100%.
To enable this programming through to the main menu, select
CONF
(ENTER),
PRG
(ENTER),
OUTPUT
(ENTER) and
STEP
(ENTER). After this
sequence, there will have the options
10%
,
20%
,
25%
and
VARIABLE
; these options
define the percentage of variation at the output for each step. The
VARIABLE
option
allows the programming of the setpoint values for each step, up to a maximum of
eleven values.
The output type must be previously configured; otherwise a
SELECT
OUTPUT FIRST
message will be displayed. In this case, press
C/CE
to go back to
the main menu and select the output type.
After completing the selection of the step variation percentage, the start and
the end values of the range in which the output will vary (
Setpoint High and Low
)
are requested.
To continue, go back to the main menu and activate the function in
EXEC
, the
output now performs the
STEP
programming, always starting from the beginning of
presys
Summary of Contents for PC-507-IS
Page 1: ...PC 507 IS Pressure Calibrator TECHNICAL MANUAL presys Instruments p r e s y s ...
Page 10: ...Page 9 2 3 Order Code p r e s y s ...
Page 13: ...Page 12 3 Operation 3 1 Parts Identification Front Panel Fig 01 Front Panel p r e s y s ...
Page 14: ...Page 13 Left side view Panel Right side view Panel Fig 02 Side Panels p r e s y s ...
Page 31: ...Page 30 a 4 wire Transmitter Calibration Fig 13 4 WIRE Transmitter Calibration p r e s y s ...