Page 7
INSTRUMENTS
LED
LED OUTPUT
INDICATION
ACTION
Steady green light
Apparatus ready for use
Use apparatus
Green flashing light
Heating underway
Wait till heating ends
Green light flashing
with reduced
frequency
ECO-mode (energy saving)
Operate the lever to return
the unit to normal
operation, otherwise
switch the unit off.
Steady red light
Sealing has finished (where
bar has been closed for 3
seconds)
Re-open sealing bar by
raising lever
Red flashing light
Sealing bar broken
Replace the sealing bar
Lever
The lever is used to operate the sealing bar. Actuating it will either engage or
disengage the bar for sealing.
Cutting tool
The cutting tool is used to operate the blade and provides a clean and easy cut
once the instruments have been heat-sealed into a sealing bag.
ON/OFF switch
The ON/OFF switch is used to switch the power on or off.