Model 9016 

Upgrade Instructions

 

 

February 2004 

 

 

 

 

 

 

 

 

 

 

 

NetScanner™ System

 

 
 

www.

PressureSystems

.com

 

 

 
 
 

Summary of Contents for 9016

Page 1: ...Model 9016 Upgrade Instructions February 2004 NetScanner System www PressureSystems com...

Page 2: ...ssureSystems com 1 Packing List When you purchase your 9016 Upgrade Kit the following parts will be included G 1 9016 Upgrade Installation Instructions G 1 Upgrade Tag G 1 PC327 analog board G 1 Assem...

Page 3: ...The following instructions must be followed to successfully upgrade the hardware and firmware in your current 9016 Intelligent Pressure Scanner to 9116 scanner capabilities Figure 1 is an exploded vi...

Page 4: ...e cleaning use a lint free applicator with acetone to remove dirt and lightly lubricate the O ring surface with Krytox provided in the maintenance kit Never use sharp objects to cut tubing from the bu...

Page 5: ...the housing All components of the pressure scanner are attached to the top plate and will lift out of the module housing when the top plate is removed See Figure 4 next page In some cases it may be e...

Page 6: ...l of the PC 206 Amplifier Multiplexer Board Use the tools and follow the general precautions previously described 1 Carefully remove the wiring harness from connector P1 of the PC 206 board See Figure...

Page 7: ...et Microprocessor A D Board Use the tools and follow the general warnings already described 1 Carefully remove any attached wiring harnesses from connectors P1 P3 and P6 of the PC 280 board This will...

Page 8: ...will be replaced by the PC 322 Mother Board and the PC 323 Microprocessor Board The PC 322 and the PC 323 will arrive attached to one another and they will be installed as a single unit Installing th...

Page 9: ...allation of the old PC 280 board See Figure 8 1 Use the same precautions as previously noted when removing these boards from their protective shipping container i e prevention of electrostatic dischar...

Page 10: ...ap be careful not to nick or cut any other wires or tubes 3 Attach the wiring harnesses to their respective connectors P1 P3 and P6 Note that the connector from P3 connects to your new PC 327 board It...

Page 11: ...les and tubing are not crimped or restricted and attach the PC 322 323 boards to the chassis with the three 3 Phillips head screws through the top plate that were previously removed and saved Do not t...

Page 12: ...the twelve 12 Phillips head screws around the top plate perimeter 3 Install the upgrade kit tag on the end of your scanner just below the serial number nameplate The tag is designed to fit directly be...

Page 13: ...Headquarters Factory Pressure Systems Inc 34 Research Drive Hampton VA 23666 USA Phone 757 865 1243 Toll Free 800 328 3665 Fax 757 865 8744 E mail sales PressureSystems com...