Accent 1000 User’s Guide-Empower
266
Operating Environment
Ambient temperature
0°C to 40°C (32°F to 104°F)
Relative humidity
10% to 90%
Atmospheric pressure
80 kPa to 106 kPa
The device needs to be protected from moisture during use.
Transport and Storage Environment
While in transport or storage, the range of temperatures to which the device can be exposed is -
25°C (-13°F) without relative humidity control to 70°C (158°F) at a relative humidity up to 93%,
non-condensing, with an atmospheric pressure range of 70 kPa to 106 kPa.
Temperature Conditions for Direct Contact
Conditions for Safe Contact – Time
Accessible Part
Contact Time Limit
Touchscreen
Less than 1 minute
Power button
Less than 1 minute
Volume button
Less than 1 minute
Conditions for Safe Contact – Temperature
Accessible Part
Maximum Temperature while Device is in
use (ambient temperature 40°C or 104°F)
Touchscreen
54.3°C (129.74°F)
Power button
50.5°C (122.9°F)
Volume button
41.2°C (106.16°F)
While the device is charging, touching the surface of the device with broken skin may
aggravate a wound.
While the device is charging, infants or high-risk groups should not touch the surface of
the device if there is a chance of burning the skin.