System Settings
Precisa Gravimetrics AG
51
Note:
The internal clock is equipped with a backup battery, which guarantees perfect functioning for
several years. If the time or dates are incorrect after a cold start, in most cases this indicates the
end of the lifespan of the backup battery. This must be replaced by qualified personnel.
6.7
Minimal Sample Weight and Quality Management,
MSW
With the minimum wine scale MWS, weighing can always be ensured in accordance with required
QS guidelines such as GLP, GMP or USP (United States Pharmacopeia).
By activating the minimum weight (MSW,) the minimum weight can be entered for several tare
ranges.
Minimum sample
weight use or not
Minimum weight i [g]
is the minimum weight that is required in tare range i
Tare range i (max) [g]
is the maximum tare value below which the minimum weight i is
used
Expiry date
when a new check of the minimum sample weight according to your QM system
is due. We certainly recommend checking once a year.
In many applications only very small quantities are used, thus, only a small part of the balance's
weighing capacity is used. The smaller the used sample quantity, the greater the relative
uncertainty of the measurement. Therefore the minimum weight which is required is determined on
the basis of the QM criteria and of the statistical data from repeated weighing procedures. The
Summary of Contents for 350-8182-000
Page 1: ...S W I S S M A D E 390 HF HA HE HM Operating Instructions 350 8182 000...
Page 11: ...390 Your new balance Precisa Gravimetrics AG 10 1 6 Assembling Optional Click Click...
Page 12: ...390 Your new balance Precisa Gravimetrics AG 11 45 Click...
Page 13: ...390 Your new balance Precisa Gravimetrics AG 12...
Page 26: ...390 Discover the Capabilities of the series 390 Precisa Gravimetrics AG 25...