Operation Manual
This brief manual is primarily meant for quick reference to wiring connections and parameter
searching. For more details on operation and application; please log on to www.ppiindia.net
LabCon Ultra
Multi-Purpose Temperature Controller
with Recording & PC Software
Operator
SP Edit on Home Screen
Sub-Level
Level
Parameters
Range
(Default)
(Default : 25.0 °C)
Temp. Setpoint Low Limit
to Temp. Setpoint High Limit
Temperature
Low Deviation Alarm
Temperature Set Value
(Default : 2.0 °C)
0.2 to 99.9 °C
Temperature
High Deviation Alarm
(Default : 2.0 °C)
0.2 to 99.9 °C
Enable
Disable
Recording
Clock & Calendar
Recording Interval
Clock Time (HH:MM:SS)
‘Delete Record’ Command
NA
0 to 250 Minutes
(Default : 5 Minutes)
Calendar Date (DD/MM/YY)
NA
NA
Sub-Level
Level
Parameters
Control Settings
Heat Zone PID Constants
Heat Zone Proportional Band
0.1 to 999.9 °C
(Default : 50.0 °C)
Heat Zone Integral Time
(Default : 100 Sec.)
0 to 3600 Sec.
Heat Zone Derivative Time
(Default : 16 Sec.)
0 to 600 Sec.
Cool Zone PID Constants
Cool Zone Proportional Band
0.1 to 999.9 °C
(Default : 50.0 °C)
Cool Zone Integral Time
(Default : 100 Sec.)
0 to 3600 Sec.
Cool Zone Derivative Time
(Default : 16 Sec.)
0 to 600 Sec.
Proportional Band
0.1 to 999.9 °C
(Default : 50.0 °C)
Integral Time
(Default : 100 Sec.)
0 to 3600 Sec.
Derivative Time
(Default : 16 Sec.)
0 to 600 Sec.
Dual Zone
Single Zone
Output Cycle Time (Sec.)
(Default : 10.0 Sec.)
0.5 to 100.0 Sec.
Single Zone & Dual Zone
Factory
Range
(Default)
Timer Start / Abort Command
Time Interval (HH:MM)
(Default : 0.10)
0.00 to 500.00
Set Value
(Temperature Control Loop)
Temp. Setpoint Low Limit to
Temp. Setpoint High Limit
(Default : 25.0)
Low Limit
(for Temp. Control Set Value)
-199.9 to Temperature
Setpoint High Limit
(Default : 10.0)
High Limit
(for Temp. Control Set Value)
For RTD : Temperature Setpoint
Low Limit to 600.0
For mA/V : Temperature Setpoint
Low Limit to 999.9
(Default : 60.0)
Self Tune
NA
Overshoot Inhibit
Enable
Disable
(Default : Disable)
Cutoff Factor
1.0 to 2.0
(Default : 1.2)
Control PV Strategy
MAP-0 PV, Average PV if
MAP-0 Fails, Average PV
(Default :
)
MAP-0 PV
NA
Zero Offset
(Default : 0.0°C)
-50.0 to 50.0°C
Compressor SP
Cool Hysteresis
Supervisory
Standby Switching
‘Switch Main/Standby Outputs’
NA
Maintenance
Repair Acknowledge
‘Control Gadget’
NA
Door Lock Access
Lock Position (On / Off)
Password Entry
Solenoid ON
Solenoid OFF
(Default : Solenoid OFF)
NA
SMS Alert
SMS
GSM Machine ID
Reset GSM Module
NA
(Default : 1)
1 to 128
Supervisory
MAP-0 Sensor
Input Type
Signal Range Low
20.00
20.00
5.000
10.00
5.000
Signal Low
to 5.000
Signal Low
to 10.00
Signal Low
to 5.000
Input Type
Settings
Default
0 to 20mA
4 to 20mA
0 to 5V
0 to 10V
1 to 5V
Signal Low
to 20.00
Signal Low
to 20.00
Input Type
Settings
Default
0 to 20mA
0.00 to
Signal High
0.00
4 to 20mA
0 to 5V
0 to 10V
1 to 5V
4.00 to
Signal High
0.000 to
Signal High
0.00 to
Signal High
1.000 to
Signal High
4.00
0.000
0.00
1.000
Signal Range High
Display Range Low
-199.9 to Range High
(Default : 0.0)
Display Range High
Range Low to 999.9
(Default : 100.0)
Zero Offset
-50.0 to 50.0°C
(Default : 0.0°C)
Alarm Settings
Inhibit
Enable, Disable
(Default : Enable)
Low Alarm Deviation
High Alarm Deviation
0.2 to 99.9
(Default : 2.0)
Hysteresis
0.1 to 99.9
(Default : 0.2)
0.0 to 100.0
(Default : 45.0)
0.1 to 99.9
(Default : 2.0)
Heat-Cool Select
Timer
Timer
Enable
Disable
(Default : Disable)
Hold Band Type
None, Up, Down, Both
(Default : None)
Hold Band Value
(Default : 0.5)
0.1 to 999.9
Start Band Value
0 to 999.9
Default : 0.5
Power Up recovery
Abort, Restart, Continuous
Default : Abort
On Timer END
Heater Off
On Timer END
Compressor Off
Yes / No
(Default : Yes)
Yes / No
(Default : Yes)
TC-J, TC-K, TC-P, TC-R,
TC-S, TC-B, TC-N,
RTD Pt100, 0 to 20 mA,
4 to 20 mA, 0 to 5 V,
0 to 10 V, 1 to 5 V
(Default : RTD Pt100)
Mode
Continuous OFF, Continuous ON,
SP Based ON-OFF, PV Based ON-OFF
(Default :
)
Continuous OFF
Time Delay
0 to 1000 Sec.
(Default : 200 Sec.)
Boundary SP
0.0 to 100.0
(Default : 45.0)
Zone Select
Single, Dual
(Default : Singlel)
Sub-Level
Level
Parameters
Range
(Default)
Configuration
Mapping
Factory
Select Mapping Inputs
1 MAP (M0),
2 MAP (M0+M1),
3 MAP (M0+M1+M2),
4 MAP (M0+M1+M2+M3),
5 MAP (M0+M1+M2+M3+M4)
(Default : 1 MAP (M0)
Input Settings
TC-J, TC-K, TC-P, TC-R,
TC-S, TC-B, TC-N,
RTD Pt100, 0 to 20 mA,
4 to 20 mA, 0 to 5 V,
0 to 10 V, 1 to 5 V
(Default : RTD Pt100)
Input Type
Signal Range Low
Input Type
Settings
Default
0 to 20mA
0.00 to
Signal High
0.00
4 to 20mA
0 to 5V
0 to 10V
1 to 5V
4.00 to
Signal High
0.000 to
Signal High
0.00 to
Signal High
1.000 to
Signal High
4.00
0.000
0.00
1.000
Signal Range High
Signal Low
to 20.00
Signal Low
to 20.00
Signal Low
to 5.000
Signal Low
to 10.00
Signal Low
to 5.000
20.00
20.00
5.000
10.00
5.000
Input Type
Settings
Default
0 to 20mA
4 to 20mA
0 to 5V
0 to 10V
1 to 5V
Display Range Low
Display Range High
-199.9 to Range High
(Default : 0.0)
Range Low to 999.9
(Default : 100.0)
Standby
Inhibit Time (Hrs.)
Fail Detect Time (Min)
0 to 250 Min.
(Default : 10 Min.)
Cyclic Time (Hrs.)
0 to 500 Hrs.
(Default : 48 Hrs.)
0 to 250 Hrs.
(Default : 1 Hrs.)
Door Open
Detection Enable
No / Yes
(Default : No)
Door Open Logic
Switch Open, Switch Close
(Default : Switch Close)
Alarm Delay (Sec.)
0 to 1000 Sec.
(Default : 60 Sec.)
Power Fail
Detection Enable
No / Yes
(Default : No)
Power Fail Logic
Switch Open, Switch Close
(Default : Switch Close)
Control Settings
Heater Output Type
(Main & Standby Heater)
SSR
0-20 mA
0-4 mA
0-5 V
0-10 V
(Default : SSR)