User Manual

Autoclave Controller

Clave

Summary of Contents for claveX 48 x 48

Page 1: ...User Manual Autoclave Controller Clave...

Page 2: ...ELLAYOUT 1 2 BASIC OPERATION 3 3 SET UPMODE ACCESS OPERATION 6 4 INPUT OUTPUT CONFIGURATION PARAMETERS 8 5 SOAK TIMER PARAMETERS OPERATION DETAILS 9 6 SUPERVISORYPARAMETERS 11 7 ELECTRICALCONNECTIONS...

Page 3: ...or the balance Soak Time value depending on the autoclave cycle state In Program Mode the Lower Readout displays parameter names prompts INDICATORS TheTable 1 1 lists each front panel LED and the asso...

Page 4: ...Pressing once decreases the value by one count keeping pressed speeds up the change PAGE User Manual Clave Table 1 1 Status Indicator HTR AIR Indicates Heater On Off status Indicates Air Outlet valve...

Page 5: ...icating that the SP value is being shown on the Lower Readout Both the heater and air outlet are maintained ON As soon as the Temperature reaches theAir Outlet SP the air outlet is turned OFF Step 2 T...

Page 6: ...e deviates above the control setpoint by more than the set band value see Operator Page and Parameters 3 PV Error The Alarm activates if the measured temperature value crosses the specified Max or Min...

Page 7: ...e AIR OUTLET SETPOINT The setpoint value at and above which the Air Outlet valve is kept closed to build saturated steam pressure The Setpoint value at which the autoclave temperature value is maintai...

Page 8: ...s to the MAIN Display Mode 4 Press and release the ENTER key until the prompt for the required parameter appears on the Lower Readout The last parameter in the list rolls back to the first parameter 5...

Page 9: ...h the Lock enabled The Figure 3 2 below illustrates the Locking procedure Figure 3 2 UnLocking Repeat the Locking procedure twice for unlocking MASTER LOCKING The controller facilitates locking all th...

Page 10: ...Control and Air Outlet setpoints HYSTERESIS Sets a differential dead band between the ON and OFF heater states Keep it large enough to avoid frequent switching of the heater without losing the desire...

Page 11: ...1 to 5 0 Default 1 0 Re Start Abort Continue Default Continue Refer end of this section for detailed Soak Timer Operation Table 5 1 TIMER START BAND The temperature band around the control setpoint T...

Page 12: ...on The timer is also provided with a Hold Band that can be enabled to make sure that the timer counts down only while the Temperature Value is within the Hold Band That is the timer pauses stops count...

Page 13: ...esired rapid changes in the measured The higher the filter Temperature Value value the better the averaging but the slower the response to actual changes OFFSET FOR TEMPERATURE VALUE This parameter ad...

Page 14: ...oth reading and writing No The Read Write parameters can only be accessed for reading That is the parameter values cannot be altered through serial communication Available for Utility Option Serial Co...

Page 15: ...Use low resistance copper conductor leads of the same gauge and length Avoid joints in the cable Section 7 User Manual Clave 48 X 48 96 X 96 OUTPUT 1 Heater Control Output The Output 1 is factory conf...

Page 16: ...ual Clave Line Neutral 12 L 13 N 2 Pole Isolating Switch Fuse Power Supply Terminal POWER SUPPLY Figure 7 6 Figure 7 4 a 18 17 16 NO C NC 16 17 18 48X48 96X96 As standard the controller accepts single...

Page 17: ...a n e t Process Precision Instruments 101 Diamond Industrial Estate Navghar Vasai Road E Dist Palghar 401 210 Maharashtra India 0250 2391722 33 37 42 07498799226 09321985369 support ppiindia net sales...