TARGAS-1 Operation Manual V. 1.02
99
Injection – Settings Menu
Syringe Volume
Button
Volume of injection syringe. Valid entry range is between 1.0 – 1,000.0 ml.
Correction Factor
Button
This user defined factor can be used to scale calculated gas concentrations to
make them match a known reference gas standard. Range: 0.50 – 2.00.
Stop Button
Stops the Custom process and returns to the Processes Menu.
Right Arrow Button
Continues the Injection process to Injection – Sample Setting screen (Step 3).
Injection – Sample Setting (Step 3)
The sample number is defined in this screen. Once the Start button is pressed, the Injection process will
begin. Before you press Start, take your syringe sample (e.g., from a chamber or airbag) if you have not
done so already. We recommend that you draw more gas from your sample source than required, and
then squeeze out excess. For example, for a 10 ml sample you should draw at least 11 ml into your
syringe, and then slowly push the plunger to the 10 ml mark.
Injection – Sample Setting Menu
Sample Number
Button
Sample number of measurement.
Stop Button
Stops the Custom process and returns to the Processes Menu.
Start Button
Starts the Injection Process and displays the Injection Zero screen (Step 4).
Injection - Zero (Step 4)
This allow the opportunity to do a zero prior to measuring sample