Gently remove the base from the upper enclosure as shown below:
Hydrophobic Filters
There are three internal hydrophobic filter located inside the TARGAS-1 enclosure to protect the
analyzers from water ingestion. They are all secured in place by black Viton tubing. These filters should
not require replacement unless you are working in extremely high humidity conditions or if you incur any
regular flow related messages. If you are regularly seeing “low flow” messages, or if you are unable to
achieve flow rates greater than 50 cc/min, then the “In-line” hydrophobic filters should be replaced. Simply
remove the filter from the tubing and replace it with a new one.
Re-Order Information
Part Number
Description
10045-1
Hydrophobic Filter
In-line
Hydrophobic
filters (3 Ea)
Battery