![PP Systems EGM-4 Operator'S Manual Download Page 35](http://html1.mh-extra.com/html/pp-systems/egm-4/egm-4_operators-manual_1592673035.webp)
Operation
EGM-4 Operator's Manual – Version 4.19
- 35 -
3CAL
To calibrate the EGM-4, press key 3.
The EGM-4 will display a value. Before proceeding, it is critical that:
•
The absorber column is properly seated in its manifold
•
The soda lime is fresh.
•
The EGM-4 has been on for at least 30 minutes.
•
The Gas Out port is unobstructed.
•
No accessories or sensors can be connected to the EGM-4 when recalibrating.
Connecting Calibration Gas To The EGM-4
During calibration, the signal from the CO
2
detector is referenced against a gas of known CO
2
concentration. The EGM-4 calibration procedure uses two reference points:
1. Zero CO
2
(generated from air that has been scrubbed through soda lime).
2. Span Gas (a user-defined CO
2
concentration, normally a cylinder of certified CO
2
composition).
It is therefore vital that both of these references are accurate. For greater accuracy, cylinder
mixtures should be accurate to +/- 1% and traceable to NIST standards. The zero gas will
only be accurate if the soda lime in the absorber column is fresh.
Assuming that a calibration gas mixture is used to calibrate the EGM-4, follow these
instructions:
1. To avoid excess pressure in the sample cell and possible damage to the Analyzer, you
must connect the sample line with a “T” piece between the calibration gas mixture and the
“Gas In” port on the EGM-4 as follows:
CO2 CONC = 2000
CORRECT (Y-N)?
EGM-4
CO
2
Gas
Cylinder
Pressure
Regulator
In-line “T” piece
to vent excess
Gas In
Port
Shut-Off
Valve
Excess gas of
approx. 250 ml/min