3 First Steps
3.4
Cabling
WARNING!
Electric shock caused by missing protective earth cable!
In the case of a fault, missing or improperly connected protective earth cables
can lead to dangerous voltages that cause an electric shock when the devices
are touched.
» Before operating the measurement system, always connect the protective
earth cables of the controller, the objective positioner (with MSA-A-TMS
option) and the sensor head as described in S
» Make sure that the contact resistance is < 0.1
Ω
at 25 A at all connection
points relevant for the assembly of the protective earth cables.
» Never remove the protective earth cables during operation.
WARNING!
Electric shock caused by dangerous voltage and residual charge on
piezo actuators!
Piezo actuators are installed both in the sensor head and in the objective
positioner (with MSA-A-TMS option). Temperature changes and compressive
stresses can induce charges in piezo actuators. If the equipment is under
voltage or residual voltage, touching the cable connections or short-circuiting
the contacts can lead to an electric shock (120 VDC max.).
» Do not remove any connection cables from the sensor head during
operation.
» Do not touch the contacts of the device plugs or connecting plugs.
» Always turn off the devices and wait some time before removing the
connecting cables.
» Secure the connection plugs on the sensor head against accidental
disconnection.
ATTENTION!
Damage to the cables caused by lack of diligence!
Carefully handle the cables so that the transmission quality is not affected
and the functionality of the devices is ensured.
» Do not expose the cables to high temperatures.
» Prevent the cables from being damaged.
» The cable bending radius must not be less than 50 mm.
» Do not expose the cables to high tensile forces.
» Do not twist the cables.
Summary of Contents for MSA-600
Page 6: ...iv Contents...
Page 20: ...1 14 1 Safety Information...
Page 24: ...2 4 2 Introduction...
Page 42: ...3 18 3 First Steps Figure 3 15 Holder with objective adapter...
Page 67: ...3 43 3 First Steps Figure 3 31 Display of sine fit in the analyzer green graph...
Page 74: ...4 4 4 Operating...
Page 80: ...5 6 5 Troubleshooting...
Page 94: ...A 2 A Declarations of Conformity Figure A 2 UKCA declaration of conformity...
Page 98: ...iv Index...