110-511 PSC/EN
8
Installation and Startup
Your MX Immersion Circulator is easily installed. The only tools required are a flat-head screwdriver and a
container for adding water or another suitable fluid to the user-supplied reservoir.
1.
Place the flow adjusters in the desired position.
These slide controls allow you to easily adjust
the force with which liquid will circulate in the reservoir. You’ll achieve maximum circulation by
placing both the front and rear flow adjusters in the fully open position.
WARNING:
To avoid the potential for burns, do not change the position of the flow adjusters
while your Immersion Circulator is immersed in hot liquid. For safety, turn power to the unit
off, remove from reservoir, and allow to cool before making adjustments.
2.
Attach the Immersion Circulator to the reservoir
(user-
supplied). A mounting clamp is provided that mounts securely to
both flat and rounded tank walls. Allow a minimum of 1.3 cm /
0.5 inch clearance between the bottom of the Circulator’s
housing and the bottom of the reservoir.
3.
Add liquid to the reservoir
. The liquid in the reservoir should
be maintained somewhere between the “Maximum” and
“Minimum” liquid level lines embossed on the heater/pump
housing.
WARNING:
See
Technical Information
in the rear of this manual for a list of compatible
liquids.
WARNING:
Read the safety data sheet for the bath fluid being used carefully before filling
reservoir.
WARNING:
If the proper fluid level is not maintained, the heater coil may become exposed
and possibly damaged (fluid level too low) or the bath may overflow (fluid level too high).
4.
Plug the power cord into a properly grounded electrical outlet
. The LCD will light and
“Standby” will appear on the display
.
WARNING:
The correct voltage and frequency for your Immersion Circulator are indicated
on the identification label on the back of the unit. The use of an extension cord is not
recommended. If one is necessary, it must be properly grounded and capable of handling
the total wattage of the unit. The extension cord must not cause more than a 10% drop in
voltage to the unit.
WARNING:
Make certain that the electrical outlet is the same voltage and frequency as
your Immersion Circulator.
Maximum
liquid level
Minimum
liquid level
www.
GlobalScientificSupply
.com
GLOBAL SCIENTIFIC SUPPLY