Instructions for Use 2019-03-20
Replacing the Built-in Electrode
• Always use the guide bar and make sure it will not come out. If the guide bar is not properly fix in place, the internal contact could be bent, and then the device might not be able to
make accurate measurements or else it might malfunction.
• Do not run the motor with the guide bar inserted; this could damage the device.
Take out the screw and then take out the built-in electrode.
Hold the File Release Button and slide the guide bar straight in as
shown in the illustration. Then rotate it left and right.
Put a little ethanol (70 vol% to 80 vol%) on a brush and clean the
rotor axle with it.
Slide the built-in electrode onto the guide bar and line up the
screw holes.
Blow air on the electrode to remove any remaining moisture.
Rotor Axle
Slowly turn the screw and make sure the built-in electrode goes
into the head properly.
Guide Bar