Factory presets for voxel size 75 micrometres
PATIENT SIZE
kV VALUE
mAs VALUE
mAs VALUE
WITH ULD
Child (XS)
90
125
25
Small adult (S)
90
125
25
Medium-sized adult (M)
100
100
20
Large adult (L)
110
90
18
Extra large adult (XL)
120
80
16
If you need to adjust the preset exposure values for this exposure:
1. Select this field to open a pop-up window.
2. Use the minus or plus signs to set the exposure values you wish to use.
To improve the image contrast, reduce the kV value. To reduce the
radiation dose, reduce the mAs value.
3. Select the green check mark button.
11.9 Selecting Artefact Removal Algorithm (ARA) and patient movement
correction (CALM)
1. Select this field to open a pop-up window.
2. Select this button to open another pop-up window.
3. Toggle the ARA (Artefact Removal Algorithm) button to select the setting
you wish to use:
•
No artefact removal
11 3D exposure