30.5.5 Suction tube cleaning holder and separation tank
Clean the suction tube cleaning holder and the separation tank behind it
once a month.
1. Open the cuspidor door.
2. Pull out the suction tube cleaning holder.
The plug can be detached from the cleaning holder prior to cleaning, but
you can just as well clean the cleaning holder with the plug still attached.
3. Disinfect the suction tube cleaning holder in a washer-disinfector at
93°C.
4. When the suction tube cleaning holder is removed you can access the
separation tank. Clean the separation tank using water and a brush. Be
careful not to dislocate the tubes in the tank.
30 Cleaning and disinfection